1 December 2008 A New Method for Calibration and Response Measurement of a Scintillation Detector Using Radioisotope Sources
Hidehito Nakamura, Hiroyasu Ejiri, Hisashi Kitamura
Author Affiliations +
Abstract

Nakamura, H., Ejiri, H. and Kitamura, H. A New Method for Calibration and Response Measurement of a Scintillation Detector Using Radioisotope Source. Radiat. Res. 170, 811–814 (2008).

A new method for calibration was developed that takes the energy deposited by radioisotope sources into account; it was validated with a plastic scintillation plate using a 207Bi source. The energy deposits of the conversion electrons in the 207Bi source led to an energy difference of a few keV for the 976 keV monoenergetic peak. Calibration with high accuracy is essential to estimate radiation doses as well as to evaluate detector performance. The method will be available not only for use in scintillation detectors but also for other detectors.

Hidehito Nakamura, Hiroyasu Ejiri, and Hisashi Kitamura "A New Method for Calibration and Response Measurement of a Scintillation Detector Using Radioisotope Sources," Radiation Research 170(6), 811-814, (1 December 2008). https://doi.org/10.1667/RR1495.1
Received: 9 June 2008; Accepted: 1 September 2008; Published: 1 December 2008